Search Thermo Fisher Scientific
- Order Status
- Quick Order
-
Don't have an account ? Create Account
Search Thermo Fisher Scientific
We are dedicated to your success. Get back on track. View our expert recommendations for commonly encountered problem scenarios.
View the relevant questions below:
Beginning your experiment?
Visit our
There are several reasons why this message may appear. Please check the following on your instrument:
If this occurs, view the raw fluorescence values for the standards and samples under “Check Standards” or "Check Calibration." Confirm that the values for the samples fall between the values of the standards (or a little above the highest standard). If they do not, the sample is out of the accurate range of the assay. The Qubit® Fluorometer screen should indicate OUT OF RANGE.
Dilute your sample, or use a smaller volume of sample in the assay and try again. Alternatively, if you are using the Qubit® DNA HS Assay or the Qubit® RNA HS Assay, try using the Qubit® DNA BR Assay or Qubit® RNA BR Assay, as these have lower sensitivity.
You can try to use a larger amount of sample if possible (up to 20 µL). Alternatively, if you are using the Qubit® DNA BR Assay or Qubit® RNA BR Assay, try using the Qubit® DNA HS Assay or Qubit® RNA HS Assay, as these have higher sensitivity.
Most likely, the sample is contaminated with other molecules that absorb at 260 nm. The NanoDrop® Spectrophotometer is reading these contaminants, but the Qubit® Fluorometer is not. To determine the exact composition of the sample, perform a Qubit® dsDNA Assay, Qubit® RNA HS Assay, and Qubit® Protein Assay on small aliquots of the same sample. Check the Qubit® assay kit product manual for a list of contaminants that could interfere with the assay. Dilute or purify the sample further to reduce contaminant concentration. Mix the working solution and the sample aliquoted into it well.
Please see our suggestions below:
Please see the possibilities below:
Standards for the 2 μL aliquots were made from 500 µg/mL lambda phage DNA stock diluted to 25, 50, 100, 300, 400, and 500 µg/mL. Standards for the 10 μL aliquots were 5X dilutions of the 2 μL standards. Standards for the 20 μL aliquots were 2X dilutions of the 10 μL standards. Note that using 2 µL of the highest concentrations of lambda DNA standards gave lower than expected fluorescence in the assay.
Here are several suggestions:
For the Qubit® 1.0 instrument, if it is connected to a computer, disconnect it from the computer, unplug it, wait at least 10 seconds, then plug it back in.
For the Qubit® 2.0 instrument, power off or unplug the instrument if possible, wait at least 10 seconds, then plug it back in. Do not unplug the instrument or remove the USB stick during a firmware update.
Please download and install the most current version of the Qubit® Firmware from our website.
For a Qubit® 2.0 Fluorometer with a serial number below 1104004846, please use a Windows® operating system to update your Qubit® 2.0 Fluorometer using the instructions on this page. If you do not have access to a PC, please contact Technical Support at techsupport@lifetech.com for further instructions.
Click on the appropriate link below to install the needed firmware update:
Qubit® 2.0 Firmware Update
Qubit® 1.0 Firmware Update (see bottom right-hand corner of the page)
Absorbance readers cannot distinguish small fragments, degraded samples, or single nucleotides from intact nucleic acid. The dyes in the kits only detect intact strands longer than about a 20-mer. The kits are only recognizing intact DNA or RNA, and ignoring degraded sample.
Negative fluorescence is a physical impossibility. It is an artifact from software autocorrecting for background signal. This means your reader is picking up and subtracting out background light at the cost of your data. Make sure to do a buffer-only control and assess the type of signal. You may need to switch to a different plate.
For Research Use Only. Not for use in diagnostic procedures.