Search Thermo Fisher Scientific
Search Thermo Fisher Scientific
Our robust ICP-OES systems offer easy-to-use software and multi-element detection technology far superior to that of single-element AAS and multi-element microwave plasma techniques. They are ideal for laboratories with low sample throughput requirements and users new to the technique or who require a simple solution for multi-elemental analysis.
Analyze high-matrix trace elemental samples with sensitive multi-element detection and meet your data requirements. These instruments include both the iFR analysis mode, which measures the entire wavelength range in one measurement, and a eUV analysis mode, which measures the UV region of the spectrum with greater sensitivity.
Do you need to screen samples for their elemental composition as part of method development or to inform further analysis? The new semi-quantitative analysis tool within the Qtegra ISDS Software leverages artificial intelligence to enable you to do this in the simplest way. Find out how by reading this product spotlight.
Are you curious to see what our iCAP PRO Series ICP-OES Systems look like close up? We've designed a 3D product tour to help you explore these systems in a self-guided manner. Find videos, literature, and detailed descriptions of key features and capabilities.
iCAP PRO ICP-OES | iCAP PRO X ICP-OES | iCAP PRO XP ICP-OES | iCAP PRO XPS ICP-OES | ||
---|---|---|---|---|---|
Plasma view | Vertical Radial or Vertical Duo | ||||
Start up time | 1 hour (from off) | 15 minutes (from standby) | 5 minutes (from standby) | ||
Wavelength range | iFR mode 167.021 nm - 852.145 nm | iFR mode 167.021 nm - 852.145 nm eUV mode 167.021 nm - 240.063 nm | |||
Resolution @ 200nm | 7 pm | ||||
Measurement modes | iFR | iFR and eUV | |||
Camera technology | CID821 | ||||
Full spectrum capture | Yes | ||||
RF source radial | 1150 W | 750, 1150 or 1350 W | 750 to 1600 W | ||
RF source duo | 1150 W | 750, 1150 or 1350 W | 750 to 1600 W | ||
Gas flow controls | Optimized MFC | Variable MFC | |||
Plasma gas | Fixed at 12.5 L/min | Fixed, optimized at 8.5, 12.5 or 14.5 L/min | 0 to 20 L/min | ||
Auxiliary gas | Fixed at 0.5 L/min | Selectable 0.5, 1 or 1.5 L/min | 0 to 2 L/min | ||
Nebulizer gas | 0.3 to 0.8 L/min | 0 to 1.5 L/min | 0.0 to 1.5 L/min | ||
Additional gas | — | 0 - 0.25 L/min |
Learn how the iSC-65 Autosampler offers exceptional performance in streamlining workflows and sample throughput.
The Thermo Scientific iSC-65 Autosampler is a workhorse partner of the iCAP PRO Series ICP-OES. It has been designed with reliability from the ground up to ensure all samples are analyzed the first time.
Reduce your lab’s environmental impact with the iCAP PRO Series ICP-OES, recipient of the ACT label from My Green Lab. The ACT label, the standard for determining the environmental impact of laboratory products, helps labs make informed, sustainable purchasing decisions.
* Required field