Search Thermo Fisher Scientific
Search Thermo Fisher Scientific
The Thermo Scientific Multiskan SkyHigh Microplate Spectrophotometer is a UV/Vis microplate spectrophotometer designed to be convenient and easy to use for virtually any photometric research application, especially DNA, RNA, and protein analysis, as well as turbidity measurements. The Multiskan SkyHigh system is offered in a variety of configurations making it flexible to suit a wide variety of needs. Touch screen models offer the flexibility to use the stand-alone instrument or in conjunction with Thermo Scientific SkanIt PC software. The Multiskan SkyHigh model operated solely via SkanIt software is ideal for users who rely on a PC for all operations. Cuvette reading capability is also offered in some models. With different options to choose from and a wide range of applications the Thermo Scientific Multiskan SkyHigh Microplate reader helps you get the most out of your lab.
In addition to the common features of all Microplate Reader systems, the Multiskan SkyHigh spectrophotometer offers:
Rapid measurement of low sample volumes
Our µDrop and µDrop Duo Plate is designed for rapid measurement of low sample volumes. You can analyze up to 32 samples simultaneously and quickly and easily wipe off the samples for serial measurements.
Thermo Fisher Connect and Microsoft OneDrive
Thermo Fisher Connect and Microsoft OneDrive cloud-based capabilities enable users to access, manage, and share data remotely, available on touchscreen models.
SkanIt software
SkanIt software can be used to control the Multiskan SkyHigh spectrophotometer and supports optimal use of the instrument’s features with a visual workflow, easy data analysis, and exporting capabilities.
Flexibility
Available in three different configurations: 1) touchscreen, 2) cuvette and touchscreen, and 3) operated only with SkanIt PC Software.
Fast operation
Full spectrum (200–1,000 nm) of a well is obtained in less than 10 seconds and a full 96-well microplate is read in 6 seconds.
|
|
||||
|
Multiskan SkyHigh | Multiskan SkyHigh with Touch Screen | Multiskan SkyHigh with Touch Screen and µDrop Duo Plate | Multiskan SkyHigh with Touch Screen and Cuvette | Multiskan SkyHigh with Touch Screen, Cuvette, and µDrop Duo Plate |
SkanIt software | ✓ |
✓ |
✓ |
✓ |
✓ |
Touch screen* |
|
✓ |
✓ |
✓ |
✓ |
Cuvette reading capability |
|
|
|
✓ |
✓ |
µDrop Duo Plate |
|
|
✓ |
|
✓ |
* Allows access to Thermo Fisher Connect and Microsoft OneDrive cloud-based capabilities. Can be connected to the local area network.
General Specifications for Multiskan SkyHigh Microplate readers
Technical specifications |
Description |
Wavelength selection |
Monochromator |
Light source |
Xenon flash lamp |
Wavelength range |
200–1,000 nm with 1 nm steps |
Readout range |
Up to 4 absorbance units |
Bandwidth |
<2.5 nm |
Linearity at 450 nm* |
0–2.5 Abs, 2% |
Accuracy at 450 nm* |
1.0% + 0.003 Abs (0–2.0 Abs), 2.0% (2.0–2.5 Abs) |
Precision at 450 nm* |
SD <0.003 Abs or CV <1.0% |
Plates |
6- to 48-, 96-, and 6-384-well plates***, Thermo Scientific µDrop Plates |
Cuvette |
Dimensions (H x W x D): 40–58 x 12.5 x 12.5 mm Beam center height: 8.5 mm Beam window: ≥2 mm |
Measurement speed (starting from A1 back to A1) |
6 sec with 96-well plate 10 sec with 384-well plate |
Plate shaking |
Linear |
Spectral scanning speed |
10 sec from 200–1,000 nm with 1 nm steps |
Incubation range |
From ambient temperature +2C to 45˚C”. |
General specifications | |
---|---|
User interfaces |
Stand-alone use: 7 in. touchscreen display PC control: Thermo Scientific SkanIt Software |
Connections |
1 USB B port for PC 1 Ethernet port 3 USB A ports for devices (USB memory device and Wi-Fi dongle) |
Main input |
100–240 V (50/60 Hz) |
Max power consumption |
<110 W |
Power save consumption** |
7 W |
Dimensions (H x W x D) |
265 x 295 x 445 mm (10.4 x 11.6 x 17.5 in.) |
Weight |
True weight: 10.6 – 11.5 kg (23.4 – 25.3 lb), depending on the model |
* Specifications are valid for 96-well flat-bottom microplates and quartz semimicro cuvettes with a 4 mm beam window.
** Models with touch screen.
***Maximum plate height with lid is 19.5 mm.