Chemiluminescent detection for western blotting is popular because it offers several advantages over other detection methods. Using chemiluminescence allows multiple exposures to be made, which enables optimization of signal to noise. The detection reagents can be removed and the entire blot reprobed to visualize another protein or to optimize detection of the first protein. A large linear response range allows detection and quantitation for a large range of protein concentrations. Most importantly, chemiluminescence yields the greatest potential sensitivity of any available detection method for western blotting. These advantages have allowed chemiluminescence to become the detection method of choice in most protein laboratories.

Explore Chemiluminescent substrates  Chemiluminescent Western Blot ProtocolDownload the Chemiluminescence western blotting technical guide



Introduction

Western blotting is a powerful technique that enables indirect detection of protein samples immobilized on a nitrocellulose or polyvinylidene fluoride (PVDF) membranes. In a conventional western blot, protein samples are first resolved by sodium dodecyl sulfate–polyacrylamide gel electrophoresis (SDS-PAGE) and then electrophoretically transferred to the membrane. Following a blocking step, the membrane is probed with a primary antibody that was raised against the antigen in question. After a subsequent washing step, the membrane is incubated with a conjugated secondary antibody that is reactive toward the primary antibody. The two most common enzymes conjugated to secondary antibodies are horseradish peroxidase (HRP) and alkaline phosphatase (AP). These enzymes catalyze the chemical reaction for generating a recordable signal in the form of light.

Chemiluminescent substrates for HRP and AP differ from other substrates in that the light detected is a transient product of the reaction that is only present while the enzyme-substrate reaction is occurring. This contrasts with chromogenic substrates that produce a stable, colored product; these colored precipitates remain on the membrane after the enzyme-substrate reaction has terminated. On a chemiluminescent western blot, the substrate is the limiting reagent in the reaction; as the substrate is exhausted, light production decreases and eventually ceases. A well-optimized procedure using proper antibody dilutions will produce a stable output of light for several hours, allowing consistent and sensitive detection of proteins.

Chemiluminescent western blot detection methods

The choice of substrate for chemiluminescent western blotting is determined by the reporter enzyme that is selected. Specifically, luminol- and acridan-based reagents are chemiluminescent HRP substrates. For chemiluminescent detection of AP, acridan- and 1,2-dioxetane-based substrates are available.

HRP luminol light producing reaction

Chemiluminescent reaction of HRP substrate- luminol. Chemiluminescence is a property of chemical reactions that emit light as a byproduct. Luminol is one of the most widely used chemiluminescent reagents. The oxidation of luminol by peroxide results in creation of an excited state product called 3-aminophthalate. This product decays to a lower energy state by releasing photons of light.

CDP-Star (AP substrate) light producing reaction

Chemiluminescent reaction of AP substrate- CDP-Star.CDP-Star is dephosphorylated by AP to yield meta-stable dioxetane phenolate anion intermediate that decomposes and emits light with a maximum intensity at a wavelength of 475 nm. Light emission occurs only during the enzyme-substrate reaction.

 HRP substrates (Horseradish peroxidase)AP substrates (Alkaline phosphatase)
Sensitivity
Femogram sensitivity
Picogram sensitivity
Signal generationImmediateGradually increases with signal maximum at ~30-60 minutes
Signal durationUp to 24 h24-96 hours
ConsiderationsCompatible with common buffers such as TBS and PBSNot compatible with phosphate buffers
When to useAntibodies or Probes conjugated to HRPAntibodies or Probes conjugated to AP

Horseradish peroxidase (HRP) catalyzes the oxidation of substrates by hydrogen peroxide, resulting in a colored or fluorescent product and the release of light as a by-product of the reaction. HRP functions optimally at a near-neutral pH and can be inhibited by cyanides, sulfides and azides. Antibody-HRP conjugates are superior to antibody-AP conjugates with respect to the specific activities of both the enzyme and antibody. In addition, its high turnover rate, good stability, low cost, and wide availability of substrates makes HRP the enzyme of choice for most applications. Because of the relatively small size of the HRP enzyme, further increases in sensitivity may be achieved by using poly-HRP conjugated secondary antibodies and may eliminate the need for using ABC-type amplification systems for some researchers. Alkaline phosphatase (AP) catalyzes the hydrolysis of phosphate groups from a substrate molecule resulting in a colored or fluorescent product and the release of light as a byproduct of the reaction. AP has optimal enzymatic activity at a basic pH (pH 8–10) and can be inhibited by cyanides, arsenate, inorganic phosphate and divalent cation chelators, such as EDTA. As a label for Western blotting, AP offers a distinct advantage over other enzymes. Because its reaction rate remains linear, detection sensitivity can be improved by simply allowing a reaction to proceed for a longer time period.

As HRP is the most popular enzyme used in western blotting it will be discussed throughout this article as our example. Several varieties of Thermo Scientific Pierce ECL and SuperSignal chemiluminescent HRP substrates are available that provide different levels of sensitivity for chemiluminescence western blotting. Refer to Table 1 to identify the most appropriate HRP chemiluminescent substrate based on the abundance of your target protein of interest, abundance of sample containing the target protein, and the level of sensitivity and type of instrumentation available for detection. The table also outlines characteristics of commonly used enhanced chemiluminescent (ECL) and SuperSignal chemiluminescent substrates.

Chemiluminescent substrates for HRP are most commonly two-component systems consisting of a stable peroxide solution and an enhanced substrate solution, often luminol-based. In most cases, to make a working solution, equal volumes of the two components are mixed together. When incubated with a blot on which HRP-conjugated antibodies (or other probes) are bound, a chemical reaction emits light at 425nm which can be captured with X-ray film and CCD camera imaging devices that detect chemiluminescence. Although X-ray film provides qualitative and semi-quantitative data and is useful to confirm the presence of target proteins, cooled CCD camera based imagers offer the advantages of quantitative analysis, rapid data capture, higher sensitivity, greater resolution and a larger dynamic range than film.

Table 1. Thermo Scientific chemiluminescent HRP substrates.

 Pierce ECL substratePierce ECL Plus substrateSuperSignal West Pico PLUS substrateSuperSignal West Dura substrateSuperSignal West Femto substrateSuperSignal West Atto substrate
Detection levelLow- to midpicogramLow-picogramLow-picogram to high-femtogramMid-femtogramLow- to midfemtogramLow femtogram- high attogram
Signal duration0.5–2 hr5 hr6–24 hr24 hr8 hr6 hr
Detection methodsX-ray film, CCD imagerX-ray film, CCD imager, fluorescence imagerX-ray film, CCD imagerX-ray film, CCD imagerX-ray film, CCD imagerX-ray film, CCD imager
Recommended primary antibody dilution*1:1,0001:1,0001:1,0001:5,0001:5,0001:5,000
Recommended secondary antibody dilution*1:25,000 - 1:200,0001:1,000 - 1:15,0001:20,000 - 1:100,0001:50,000 - 1:250,0001:100,000 - 1:500,0001:100,000 - 1:250,000
Select when ...... target and sample are abundant... target is less abundant, sample is limited, and you need chemifluorescent detection... target is less abundant, sample is limited, and you need more sensitivity than an entry-level ECL substrate provides... target is less abundant, sample is limited, and you are using CCD image capture... target is least abundant, sample is precious, and you need maximum sensitivity... target is least abundant, sample is precision, and you need maximum sensitivity
Value to youLow cost; easy to switch from other entry-level ECL substratesBest detection flexibility with chemifluorescent detection optionBest value; works for majority of western blotsBest signal durationGood sensitivity with optimized conditionsMost sensitivity with less optimization required
*Based on a 1mg/mL antibody concentration

Light output generated during a chemiluminescent luminol reaction is relatively short lived; therefore, enhanced chemiluminescent (ECL) substrates have been developed. For example, use of an enhancer with luminol increases signal sensitivity, intensity, and duration of the enzyme-substrate reaction. Pierce ECL substrate is appropriate for western blot applications in which abundant proteins are being probed or where the experiment has been optimized. However, substrates providing highly sensitive protein detection have been developed. For example, SuperSignal West Pico PLUS Chemiluminescent Substrate enables picogram- to high femtogram–level protein detection by western blot analysis.

Learn more:Detecting low abundant proteins in a western blot
Explore:Chemiluminescent substrate selection guide

Technical considerations

Antibody dilution

One of the most impactful technical factors in western blotting is optimizing the antibody dilutions. Many factors may influence the desired antibody dilution including volume of antibody available, antibody specificity for the antigen, protein abundance and choice of available substrates. Furthermore, individual lab protocols may direct researchers towards preferred, previously optimized conditions. Optimizing the antibody is a critical step to ensure proper signal to noise. The primary antibody, which is specific to the target antigen, binds to the available epitope sites on the protein. After a series of washing steps which helps remove unbound primary antibody, a secondary antibody, conjugated to an enzyme such as HRP, is incubated and binds to the primary antibody. Multiple secondary antibodies can bind to a single primary antibody- resulting in an amplification process whereby more light-producing labels can react and produce higher detectable signal. Too much primary or secondary antibody can cause oversaturation of the blot or excessive background noise. Also, too much antibody can lead to rapid signal formation followed by degradation – its possible for the signal to fade quickly before one could capture the signal on film or with an imaging system. These problems make it extremely difficult to properly analyze and make conclusions from the experiment. The table below can help provide general guidance on antibody dilutions depending on the protein target or sample abundance, antibody conditions, and substrate of choice. It is also important to note that when switching from one substrate to another, one may need to reoptimize their antibody dilutions based on the vendor’s recommendations.

Antibody Dilution Calculator

Antibody dilution recommendations

Conditions 
SampleAntigenAntibodyRecommended substratePrimarySecondarySpecial notes
Sample is abundantAntigen is abundantAntibody quality is goodSuperSignal West Pico PLUS1:1K1:100K 
Sample is abundantAntigen is abundant is unknownAntibody quality is unknownSuperSignal West Pico PLUS1:1K1:100KIf no bands are seen, the blot can be washed and re-imaged with Atto
Sample is abundantAntigen abundance is lowAntibody quality is poorSuperSignal West Atto substrate1:1K1:100K 
Sample is abundantAntigen abundance is lowAntibody quality is goodSuperSignal West Atto substrate1:2K1:250KAssumes normal primary usage is 1:1K
Sample is abundantAntigen abundance is lowAntibody is limitedSuperSignal West Atto substrate1:5K1:100KAssumes normal primary usage is 1:1K
Sample is abundantAntigen is abundantAntibody is limitedSuperSignal West Atto substrate1:20K1:250KAssumes normal primary usage is 1:1K
Sample is limitedAntigen is abundantAntibody quality is goodSuperSignal West Pico PLUS1:1K1:100KAlternatively, could load less and use Atto at same antibody concentrations
Sample is limitedAntigen abundance is unknownAntibody quality is unknownSuperSignal West Pico PLUS1:1K1:100KIf no bands are seen, the blot can be washed and re-imaged with Atto
Sample is limitedAntigen abundance is lowAntibody quality is goodSuperSignal West Atto substrate1:1K1:250K 
Sample is limitedAntigen abundance is lowAntibody quality is poorSuperSignal West Atto substrate1:1K1:100K 
*Based on a 1mg/mL antibody concentration

Signal capture

Western blotting is a powerful, routine application; however, capturing a chemiluminescent signal can be challenging since western blotting comprises a series of steps that require specific skills to perform. Failure to capture signal can be caused by many factors. With so many variables (Table 2), troubleshooting a problem blot can be likened to searching for a “needle in a haystack”. The traditional protocol is often ineffective in detecting a specific protein. Therefore, troubleshooting guides are frequently needed to help optimize results. For an overview of factors affecting western blot results review Table 2. To obtain western blot troubleshooting information visit, Western Blot — Troubleshooting.

Table 2. Factors that affect western blotting results.

FactorVariable characteristicQuick Tips
Target antigenConformation, stability, available epitope(s) 
Polyacrylamide gelManufacturer, percent polyacrylamide, age, lotUse Bis-Tris gels for low/medium abundant proteins, use Tris-glycine gels for high abundant proteins, Tricine gels for low molecular weight proteins, and Tris-Acetate gels for high molecular weight proteins
MembraneManufacturer, type, lot 
Primary antibodySpecificity, titer, affinity, incubation time and temperatureUse antibodies validated specially for western blotting
HRP conjugateEnzyme activation level and activity, source animal, concentration, detergent 
Blocking bufferType, concentration, cross-reactivity 
WashesBuffer, volume, duration, frequencyUse minimum of 6x 5 minutes washes prior to substrate addition to thoroughly remove unbound secondary antibody
SubstrateSensitivity, manufacturer lot, age 
Detection methodFilm age, imaging instrument manufacturer, exposure time 

Signal intensity and duration

Under optimal western blotting conditions, a chemiluminescent signal can last for 6–24 hr. The level and duration of light generation depends on the specific substrate being used and the enzyme-to-substrate ratio in the system. Although the amount of substrate on a blot is relatively constant, the amount of enzyme present depends on the amount added and other factors (Table 2). Too much enzyme conjugate applied to a western blot system is one of the biggest causes of signal variability, dark background, short signal duration, and low sensitivity. A signal emission curve that decays slowly (see figure below) is desirable as it demonstrates that each component of the system has been optimized and allows reproducible results. A signal that decays too quickly can cause variability, low sensitivity, high background, and problems with signal documentation. A long-lasting signal minimizes variability in results due to transfer efficiency, different manufacturer lots of substrate, and other factors. The oxidation reaction of the HRP molecules with luminol in the substrate produces free radicals in addition to the light being produced. An abundance of HRP in the system will create an abundance of free radicals, speeding the probability of HRP inactivation. Free radicals may also damage the antigen, antibodies, and the membrane, resulting in reduced effectiveness of re-probing. The figure below provides an example of signal emission curves generated with short- and long- duration substrates.

Comparison of signal emission for short and long duration substrates for western blotting

A comparison of signal emission curves for short- and long duration substrates. When there is an enzyme present in a western blot system, signal output peaks soon after substrate application and rapidly exhausts the substrate (Signal 1). In an optimal system, the signal emission peaks approximately 5 min after applying the substrate and plateaus for several hours (Signal 2).

Direct and indirect methods

Direct detection uses a labeled primary antibody. Because incubation with a secondary antibody is eliminated, this strategy takes less time than a classic western blot. Additionally, background signal from cross-reactivity with the secondary antibody is eliminated. Direct detection also enables probing for multiple targets simultaneously. Labeling a primary antibody, however, sometimes has an adverse effect on its immunoreactivity, and allows for no signal amplification. As the direct method is generally less sensitive than an indirect detection, it is most useful when the target is relatively abundant. One option is biotinylating the primary antibody, which is an indirect method that both amplifies the signal and eliminates the need for a secondary antibody. Labeling with biotinylation reagents typically results in more than one biotin moiety per antibody molecule. Each biotin moiety is capable of interacting with enzyme-conjugated avidin, streptavidin, or Thermo Scientific NeutrAvidin Protein (Cat. No. 31000). Essentially, the biotin-binding conjugate replaces a secondary antibody, and its appropriate molar concentration is the same as if a secondary antibody were used. The small size of biotinylation reagents (typically less than 2 kDa) compared to enzymes (40 kDa for HRP; 140 kDa for AP) often reduces steric hindrance, allowing greater functionality of the labeled primary antibody.

Direct and indirect protein detection and signal amplification by avidin-biotin complex formation in an assay involving detection with specific antibodies (e.g., an immunoassay).

Far-western methods

Occasionally, an antibody to a specific antigen is unavailable or unsuitable for western blot analysis. Target protein–specific detection by blotting is possible if a corresponding binding partner is available for use as a probe. This type of application, referred to as a far-western blot, is routinely used for the discovery or confirmation of a protein–protein interaction. There are numerous variations in western blotting detection that include, but are not limited to, the previously mentioned strategies. In far-western blotting, labeled binding partners may be labeled by an in vitro translation reaction with 35S-labeled probe. Biotinylating the probe and detecting with a biotin-binding protein conjugate is also possible. This type of detection has the added advantage of signal amplification. However, when using this technique the probe should not be over-labeled, to prevent target detection from being compromised. Additionally, a recombinant probe can be expressed in bacteria with a tag such as GST, HA, c-Myc, or FLAG, in which detection occurs via a labeled antibody to the particular tag. As with other blotting applications, the far-western method is effective for both membrane and in-gel detection systems. The diagram below outlines with far-western workflow.

Diagram of Far western blot technique to analyze protein-protein interactions

Diagram of far–western blot to analyze protein–protein interactions. In this example, a tagged bait protein is used to probe either the transfer membrane or a gel for the prey protein. Once bound, enzyme (horseradish peroxidase; HRP)-conjugated antibody that targets the bait tag is used to label the interaction, which is then detected by enzymatic chemiluminescence. This general approach can be adjusted, as shown in the table below, by using untagged bait protein that is detected by antibody, biotinylated bait protein that is detected by enzyme-conjugated streptavidin, or radiolabeled bait protein that is detected by exposure to film.

Learn more:Far Western Blot Analysis
Explore:Far-western blot kit

Data capture

Chemiluminescent western blot signal can be captured with X-ray film, charge-coupled device (CCD) camera–based digital imaging instruments, and phosphorimagers that detect chemiluminescence. Although X-ray film provides qualitative and semi-quantitative data and is useful to confirm the presence of target proteins, CCD camera–based imaging instruments offer the advantages of qualitative analysis, instant image capture and analysis, higher sensitivity, greater resolution and a larger dynamic range than film. Additionally, there is no need for dedicated darkroom space and developer equipment. Imaging instruments can be placed directly on a bench alongside other lab equipment.

CCD cameras for chemiluminescent image capture

iBright 1500 Imaging System
AdvantagesLimitations
Dynamic range with >4 orders of magnitudeExtended exposures (>60 minutes) can result in increased background from camera noise
Smart algorithms help determine optimal exposure time 
Streamlined image analysis and storage 
Instant visualization of results 
Quantitative 
iBright Imaging systems feature a 9.1 MP cooled CCD camera that provides high sensitivity and dynamic range to help enable the detection of subtle differences in samples.

Often, it is necessary to expose several films for different time periods to obtain the proper balance between signal and background. The goal is to time the exposure of the membranes to the film so that the desired signal is clearly visible while the background remains low. This is difficult to accomplish since the process cannot be observed and stopped when the desired endpoint is reached. If the film is not exposed long enough (underexposed), the signal will not be visible. If the film is exposed too long (overexposed), the signal may be lost in the background or separate bands may become blurred together. Furthermore, judging the optimal exposure time is highly subjective, as its difficult to visually determine the precise point at which the best balance of signal to noise is achieved. Repeated bouts of exposing the film and developing the film are often performed, which can add up to a considerable investment of time. After the data is captured on film, the film is often scanned to make a digital copy of the data, which is then exported to analysis software to facilitate further analysis, such as densitometry and molecular weight estimation.

One of the biggest benefits of digital imaging systems is the ability to fine-tune the exposure time (some systems offer the ability to adjust exposure time to the thousandth of a second). In addition, many systems offer specific auto-exposure setting which employ algorithms that automatically determine the optimal exposure time, which not only captures an image with optimal signal-to-noise, but also saves a considerable amount of time versus the multiple exposures often required with film. Since the data is captured digitally, it can be easily exported to analysis software (eliminating the scanning step with film). Some of the most modern systems offer the ability to perform analysis directly on the instrument. These conveniences are making digital imaging a popular replacement for film-based data capture.

Learn more:X-ray film vs digital imaging instruments
Explore:Western blot imaging systems

Molecular weight markers

For detection of any western blot, it is desirable to use prestained molecular weight markers that are transferred to the membrane along with the protein sample. The appearance of the molecular weight markers on the membrane allows estimation of molecular weights for any protein bands that are detected as well as effective separation of the proteins of interest in the gel prior to the transfer step. When chemiluminescent detection is used for western blotting, protein bands are detected on film or with digital imaging equipment. Unless modified, molecular weight markers do not show up on film, since they do not produce an output of light. For this, there are several solutions described below.

SuperSignal molecular weight protein ladder for western blot analysis

Use of SuperSignal molecular weight marker for chemiluminescent western blot detection. Western blot analysis of HA-epitope tag was performed using samples from Thermo Scientific Pierce HA-Tag IP/Co-IP Kit with Thermo Scientific SuperSignal Molecular Weight Protein Ladder in Lane 1, an HA-tagged positive control lysate in Lane 2, unbound flow-through in Lanes 3 and 4, and two elutions from the immobilized anti-HA resin in Lanes 5 and 6. The membrane was probed with an HA Tag Monoclonal Antibody (Cat. No. 26183) at a dilution of 0.1 ug/mL followed by chemiluminescent detection with Thermo Scientific SuperSignal West Dura Extended Duration Substrate (Cat. No. 34075).

iBright Prestained Protein Ladder

iBright Prestained Protein Ladder contains twelve recombinant proteins, ten (11 to 250 kDa) that are blue-stained and fluor-labeled for direct and near-IR fluorescent visualization and protein sizing, and two (30 kDa and 80 kDa) that are unstained and contain IgG binding sites to bind the primary and secondary antibodies used for chemiluminescent and fluorescent detection of the target protein.

One solution is the use of molecular weight markers that employ antibody binding domains from Protein A or G. The antibody capture domains of these proteins are engineered into the molecular weight markers and thus bind to the antibodies used in the western blot, allowing a signal to be generated and captured alongside the experimental data. The application of these markers is limited because the variable affinities of antibodies for Protein A and G result in variable levels of signal. Many antibodies, such as those of the mouse IgG1 subclass, do not bind strongly to Protein A or G. Another popular option is to use biotinylated protein molecular weight markers, which allow the use of streptavidin alongside the detection antibodies or when biotinylated primary detection antibodies are used.

Explore:Western blot protein ladders

Suggested reading

  1. G.H. Thorpe, L.J. Kricka. (1986) Enhanced chemiluminescent reactions catalyzed by horseradish peroxidase.Methods in Enzymology 133:331–54.
  2. Alegria-Schaffer, A., et al. (2009) Performing and optimizing Western blots with an emphasis on chemiluminescent detection.Methods Enzymol 463:573–99.
  3. Gassmann, M., Granacher, B., Rohde, B. and Vogel, J. (2009) Quantifying Western blots: Pitfalls of densitometry.Electrophoresis 30:1845–55.

Additional resources

For Research Use Only. Not for use in diagnostic procedures.