Software Update (Rev 2.3.1295.203)

Announcing EVOS M7000 Software Revision 2.3.1295.203

Revision 2.3.1295.203 provides a bugfix for the microscope not recognizing the following light cubes:

  • EVOS Light Cube, TagBFP 2.0, catalog number AMEP4968
  • EVOS Light Cube, QDOT WB 525-800 2.0, catalog AMEP4966
  • EVOS Light Cube, CFP-YFP EM 2.0, catalog number AMEP4969
  • EVOS Light Cube, AO 2.0, catalog number AMEP4970

It also fixes a bug that would crash the software when one of these light cubes was installed and the Settings > Light Cubes tab was clicked.

This new release also fixes a bug where, when using the OSI-2 onstage incubator, the Settings > Incubator selections for activating Port2 and Port3 were logically reversed. (Note that both gases, CO2 and N2, are still controlled correctly in the software; this bug did not switch the assigned gases in the User Interface.)

Please review the Installation Instructions below before updating your software.


Installation Instructions

Software update (Rev 2.3.1295.203)

IMPORTANT! Restart your system prior to updating to optimize the update process.

Update from USB

Uninstall the current software

  1. Go to Windows Start and select Settings, then click on Apps
  2. Under Apps and Features select M7000 from the list of installed applications
  3. Click “Uninstall” to uninstall the software

Download and Install the Software Update

  1. Insert a fresh USB drive with at least 300 MB available space.
  2. Right-click the “Download” link for Software Update.
  3. Select “Save Link as”.
  4. Select the USB Drive and press Save.
  5. Plug the USB flash drive into the M7000 computer
  6. Locate the update file from your USB drive in Windows Explorer and copy it to the hard drive.
  7. Double-click the update file from the hard drive to install the update.
  8. A Windows™ dialog box will pop up. Click Next to proceed with the software update.
  9. The screen will display the update progress. When the update is complete, click OK.
  10. Shut down the computer (Windows Start -> Power icon -> Shut down).
  11. Power up the computer and launch the M7000 software.

Recalibrate Your Instrument

  1. Open the Settings Tab.
  2. Click on Objectives, recalibrate your objectives, then click Done at the bottom.
  3. Click on Stage Calibration, recalibrate the stage, then click Done at the bottom.
  4. Recalibrate any vessels by mounting them on the stage, click Vessels, select Calibrate Vessel, follow the instructions, and when finished click Done at the bottom.

IMPORTANT! If you experience any issues with the installation or find that it is not completing as expected, please reboot your instrument and contact Technical Support.

The number for your local Technical Support Team can be found by navigating to our home page.

For Research Use Only. Not for use in diagnostic procedures.