Search Thermo Fisher Scientific
- Order Status
- Quick Order
-
Don't have an account ? Create Account
Search Thermo Fisher Scientific
We are dedicated to your success. View our expert recommendations for commonly encountered inquiries.
View the relevant questions below:
Beginning your experiment? Visit
Component | Procedure | Recommended Frequency |
---|---|---|
Inlet components | Flush (clean) the path from the sample transfer line to the source | Daily |
Housing Adapter | Clean the inner and outer electrodes | As needed |
Clean the drain adapter | As needed | |
Clean the entrance plate | As needed | |
Replace the entrance plate O-ring | As needed | |
Main control box | Replace one or both fuses | As needed |
RF transformer box | Clean the air filter | Every 4 months |
FAIMS Pro significantly improves the signal to noise ratio but with an expected signal drop off. Magnitude is sample dependent. Increasing sample loaded on column up to 2x may improve performance.
For nano-spray applications, the user defined FAIMS gas should be set to zero, with static condition.
If the DV Tune is not passing, perform the following steps:
With FAIMS attached, the spray voltage may need to be slightly higher compared to when FAIMS is not installed. This is because the FAIMS entrance plate voltage is set to 250 V. In order to keep the delta between spray tip and entrance plate similar with and without FAIMS, the voltage may need to be increased by 100-300 Volts when FAIMS is attached.
For EasySpray columns, start by using 1.8 – 2.0 kV without FAIMS, and 1.9 – 2.1 kV with FAIMS. Note that the maximum recommended spray voltage for EasySpray columns is 2.2 kV.