We offer wide variety of pre-cast gels. These include gels for analysis nucleic acids (TBE, TBE-Urea, and DNA Retardation). General information on Novex® Pre-Cast Gels is provided in this section. Novex® Pre-Cast Gels are capable of resolving proteins in the range of 2-500 kDa and nucleic acids in the range of 10-3000 bp.
Choosing a Gel for Your Application
To obtain the best results for your application, it is important to choose the correct gel percentage, buffer system, gel format, and thickness. Many factors affect the choice of a gel. These include:
Application
Based on the type of your application, you can choose from gels for protein separation (Tris-Glycine, Tricine, IEF, ZOOM® and Zymogram Gels) or gels for nucleic acid separation (TBE, TBE-Urea, and DNA Retardation Gels).
Size of the molecule being separated
Large molecules resolve well on a low percentage gels while small molecules are best resolved on high percentage gels. The size of the molecule usually dictates the acrylamide percentage. If you do not know the molecular weight of the molecule or are separating a wide molecular weight range of molecules, choose gradient gels.
Amount of available material
The higher the number of wells and the thinner the gel, the lower the sample loading volume and vice versa. Based on the amount of your starting material available, you can choose from a variety of comb types.
Compatibility
The size of a Novex® Pre-Cast Gel is 10 x 10 cm (gel size is 8 x 8 cm). We recommend using the XCell SureLock™ Mini-Cell for the electrophoresis of Novex® Pre-Cast Gels to obtain optimal and consistent performance. Novex® Pre-Cast Gels are compatible with most other mini-cells designed for electrophoresis of 10 cm (h) x 10 cm (w) gel cassettes.
The Novex® Pre-Cast Gels are compatible with most silver staining protocols. We recommend using the SilverQuest™ Silver Staining Kit or the SilverXpress® Silver Staining Kit for silver staining of Novex® Gels. The Novex® Pre-Cast Gels are compatible with any of the standard Coomassie® staining procedures. The protocols that are accelerated by heat are preferable as heat serves as a “fix” for proteins, especially smaller peptides. The SimplyBlue™ SafeStain and Novex® Colloidal Coomassie® Blue Staining Kit are recommended for staining Novex® Gels.
Applications
Performing Nucleic Acid Analysis
The Novex® TBE Gels are used to analyze DNA fragments including restriction digest, PCR products, Southern analysis, and primer analysis. The Novex® TBE-Urea Gels are used for denaturing nucleic acid analysis and are suited for RNase Protection Assays, in-vitro transcription studies, RNA stability studies, and oligonucleotide purification.
Performing Gel Shift Assays
The Novex® 6% DNA Retardation Gels are used to perform gel shift assays.
Instructions are provided below for electrophoresis of the Novex® Pre-Cast Gels using the XCell SureLock™ Mini-Cell. If you are using any other mini-cell for electrophoresis, refer to the manufacturer’s recommendations.
Preparing Samples
The Novex® Hi-Density TBE Sample Buffer (5X) can be ordered from the above table.
Reagent | Amount |
---|---|
Sample | x µl |
Novex® Hi-Density TBE Sample Buffer (5X) | 2 µl |
Deionized Water | to 8 µl |
Total Volume | 10 µl |
Preparing Running Buffer
Novex® TBE Running Buffer (5X) is available from Thermo Fisher Scientific.
Novex® TBE Running Buffer (5X) | 200 ml |
Deionized Water | 800 ml |
Total Volume | 1000 ml |
Electrophoresis Conditions
Migration of the Dye Fronts: The size of the DNA fragments visualized at the dye fronts of the different TBE Gels is shown in the table below.
Gel Type | Dye Front* | |
Bromophenol Blue (dark blue) | Xylene Cyanol (blue green) | |
6% TBE Gel
|
65 bp
|
250 bp
|
8% TBE Gel
|
25 bp
|
220 bp
|
10% TBE Gel
|
35 bp
|
120 bp
|
20% TBE Gel
|
15 bp
|
50 bp
|
4-12% TBE Gel
|
35 bp
|
400 bp
|
4-20% TBE Gel
|
25 bp
|
300 bp
|
*accuracy is + 5 bp
Protocol using XCell SureLock™ Mini-Cell
Wear gloves and safety glasses when handling gels.
XCell SureLock™ Mini-Cell requires 200 ml for the Upper Buffer Chamber and 600 ml for the Lower Buffer Chamber.
Electrophoresis Conditions
Run your gels according to the following protocol:
Gel Type | Voltage | Expected Current* | Run Time |
TBE Gels
|
200 V constant**
|
Start: 10-18 mA
End: 4-6 mA
|
30-90 minutes (dependent on gel type)
Run the gel until the bromophenol blue tracking dye reaches the bottom of the gel.
|
6% TBE-Urea Gels
|
180 V constant**
|
Start: 19 mA
End: 14 mA
|
50 minutes
Run the gel until the bromophenol blue tracking dye reaches the bottom of the gel.
|
10% TBE-Urea Gels
|
180 V constant**
|
Start: 15 mA
End: 8 mA
|
60 minutes
Run the gel until the bromophenol blue tracking dye reaches the bottom of the gel.
|
15 % TBE-Urea Gels
|
180 V constant**
|
Start: 13 mA
End: 6 mA
|
75 minutes
Run the gel until the bromophenol blue tracking dye reaches the bottom of the gel.
|
DNA Retardation Gels
|
100 V constant
|
Start: 12-15 mA
End: 6-15 mA
|
90 minutes
Run the gel until the bromophenol blue tracking dye reaches the bottom of the gel.
|
*Expected start and end current values are stated for single gels.
**Voltages up to 250 V may be used to reduce the run time.
Removing the Gel after Electrophoresis
Problem | Cause | Solution |
Run taking longer time
|
Running buffer too dilute
|
Make fresh running buffer as described in this manual and avoid adjusting the pH of the 1X running buffer.
|
Low or no current during the run
|
Incomplete circuit
|
|
Faint shadow or “ghost” band below the expected protein band
|
Ghost bands are caused due to a slight lifting of the gel from the cassette resulting in trickling of some sample beyond its normal migration point. Gel lifting off the cassette is caused due to:
|
|
Streaking of proteins
|
|
|
|
| |
|
| |
|
| |
Bands in the outer lane of the gel are curving upwards
|
|
|
|
| |
|
| |
Bands in the outside lanes of the gel “smiling”
|
Expired gels used causing the acrylamide to break down in the gel
|
Avoid using gels after the expiration date. Use fresh gels.
|
Bands are running as U shape rather than a flat band
|
Samples are loaded on the gel and not electrophoresed immediately resulting in sample diffusion
|
Load samples on to the gel immediately before electrophoresis.
|
Bands appear to be “funneling” or getting narrower as they progress down the gel
|
Proteins are over-reduced causing the proteins to be negatively charged and repel each other.
|
Reduce the proteins using DTT or b-mercaptoethanol as described.
|
Dumbbell shaped bands after electrophoresis
|
Loading a large volume of sample causing incomplete stacking of the entire sample. This effect is intensified for larger proteins
|
Load the appropriate volume of sample per well as described. If your sample is too dilute, concentrate the sample using salt precipitation or ultrafiltration.
|
For TBE-Urea gels
High background and smeared bands or abnormal band shapes
|
|
|
|
| |
|
| |
|
|