Search Thermo Fisher Scientific
- Contact Us
- Quick Order
-
Don't have an account ? Create Account
Search Thermo Fisher Scientific
Instrument calibration
To ensure optimal performance of your thermal cycler or sequence detection system, Applied Biosystems strongly recommends that you regularly calibrate the system. Calibration schedules vary by instrument; refer to the table below for the types of calibration and appropriate intervals.
Instrument calibration intervals
Calibration Type | Instrument | Interval |
ROI (Spatial) | 7300/7500/7500 Fast Real-time PCR Systems StepOne™ / StepOnePlus™ Systems | Every 6 monthsorAs needed to verify the instrument’s performance |
Background | 7900HT Fast Real-Time PCR System 7300/7500/7500 Fast Real-time PCR Systems | Every monthorBefore performing a pure dye calibration orWhen installing an uncalibrated block |
Pure dye spectra | 7900HT Fast Real-Time PCR System 7300/7500/7500 Fast Real-time PCR Systems | Every 6 months orWhen installing an uncalibrated block |
Instrument verification | 7900HT Fast Real-Time PCR System 7300/7500/7500 Fast Real-time PCR Systems | Every 6 months orAs needed to verify the instrument’s performance |
For Research Use Only. Not for use in diagnostic procedures.