Choose the best option, pipette to tip

Choose the best option, from pipette to tip

Both pipette types have a piston that moves in a cylinder or capillary. In air displacement pipettes, a certain volume of air remains between the piston and the liquid. In positive displacement pipetting, the piston is in direct contact with the liquid.

Complete the form and download a pdf version of the GLP Guide.

Download now


Air displacement pipettes

Air displacement pipettes

Air displacement pipetting is highly accurate for standard pipetting applications. However, temperature and atmospheric pressure, as well as the specific gravity and viscosity of the solution, may affect the performance of air displacement pipettes.

Aspirating the liquid (steps 1–3)
Aspirating liquid
  1. The piston moves to the appropriate position when the volume is set.
  2. When the operating button is pressed to the first stop, the piston expels the same volume of air as indicated on the volume setting.
  3. After immersing the tip into the liquid, the operating button is released. This creates a partial vacuum, and the specified volume of liquid is aspirated into the tip.
Dispensing the liquid (step 4)
dispensing the liquid
  1. When the operating button is pressed to the first stop again, the air dispenses the liquid. To empty the tip completely, the operating button is pressed to the second stop (blow-out).

Positive displacement

Positive displacement pipettes:

Positive displacement pipetting is based on direct contact of the piston with the liquid. The aspirated liquid amount depends on the dimensions of the cylinder or capillary and the movement distance of the piston. In positive displacement pipettes the tips contain both the cylinder/capillary and the piston.

piston
  1. The piston moves down inside the tip to make direct contact with the sample
piston moves up and down
  1. The piston moves up to draw the sample into the tip.
piston descends
  1. When dispensing, the piston descends and the selected volume is dispensed.

Featured product

F1 ClipTip

F1-ClipTip Multichannel Pipettes

ClipTip technology locks tips firmly in place and ensures a complete seal on every channel with minimal tip attachment and ejection forces using Thermo Scientific™ F1-ClipTip™ Multichannel Pipettes. Newly enhanced volume adjustment allows for a silent, effortless and ergonomic user experience.

Learn more

Navigate this guide:

Liquid handling
Proper pipetting techniques
Ergonomics
Maintenance & service

California Proposition 65 Warning: Products manufactured with polycarbonate (PC), polyethylene terephthalate (PET), polyethylene terephthalate glycol (PETG) or polystyrene (PS) contain chemicals known to the State of California to cause cancer, birth defects or other reproductive harm.