Berthold Technologies logo

Berthold Technologies Mithras² LB 943 and Tristar² S LB 942 have been rigorously tested using either the Terbium Assay Application Note or the Europium Assay Application Note with demonstrated performance as shown in the setup guide.

Go to Berthold Technologies website 

Before using LanthaScreen Terbium and Europium Assays

Test your plate reader setup

These technical notes provide a method for verifying that a fluorescent plate reader is able to detect a change in time-resolved fluorescence energy transfer (TR-FRET) signal, confirming proper instrument set-up and a suitable response. The method is independent of any biological reaction or equilibrium and uses reagents that are on-hand for the LanthaScreen assay.

Assay technologyAdapta and LanthaScreen EuGeneBLAzerLanthaScreen TbPredictorZ'-LYTE
ModalityTR-FRETFRET (bottom)TR-FRETFPFRET
Setup guideMithras2 LB 943 Microplate Reader Europium (Eu) Assay Setup GuideCompatible, not tested Mithras2 LB 943 Microplate Reader Terbium (Tb) Assay Setup Guide

Compatible

Contact Berthold

Tested
Excitation filter340x26405x10320x40 405x10
Emission 1 filter665xm7uv460m25520xm10uv 460m25
Emission 2 filter620xm10uv535m25495xm10uv 535m25
Dichronic mirrorFixedFixedFixed Fixed
Note: filters for the Mithras² LB 943 use the notation: excitation filters as center wavelength x bandwidth (nm) and emission filters as center wavelength xm bandwidth or center wavelength m bandwidth (nm). Contact Berthold Technologies for information on ordering filters for the microplate readers listed above.
Assay technologyAdapta and LanthaScreen EuGeneBLAzerLanthaScreen TbPredictorZ'-LYTE
ModalityTR-FRETFRET (bottom)TR-FRETFPFRET
Setup guideTristar2 S LB 942 Microplate Reader Europium (Eu) Assay Setup Guide
Compatible, not testedTristar2 S LB 942 Microplate Reader Terbium (Tb) Assay Setup Guide

Not compatible

Compatible, not tested 
Excitation filter340x26405/10320x40 405/10
Emission 1 filter665xm7uv460/25520xm10uv 460/25
Emission 2 filter620xm10uv535/25495xm10uv 535/25
Dichronic mirrorFixedFixedFixed Fixed
Excitation and Emission filters for the Tristar² S LB 942 use the notation: center wavelength/bandwidth (nm). Contact Berthold Technologies for information on ordering filters for the microplate readers listed above.

For Research Use Only. Not for use in diagnostic procedures.