Search Thermo Fisher Scientific
- Contáctenos
- Orden Rápida
-
¿No tiene una cuenta? Crear una cuenta
Search Thermo Fisher Scientific
Possible Cause | Solution |
---|---|
Reagents not at room temperature at start of assay | It is recommended that all reagents be at room temperature before starting the assay. Allow reagents to sit on bench for 15–20 minutes to reach room temperature. |
Incorrect storage of components | Double check storage conditions on kit label. Most kits need to be stored at 2–8oC. |
Expired reagents | Confirm expiration dates on all reagents. Do not use reagents that are past the expiration date. |
Reagents added/prepared incorrectly | Check protocol, ensure reagents were added in the proper order and prepared to correct dilution. |
Incorrect dilutions prepared | Check pipetting technique—see below—and double check calculations. |
Capture antibody didn’t bind to plate |
|
Not enough detector antibody used | Manufactured kits have optimized protocols. Make sure to follow recommended antibody dilutions. If developing ELISA using an Antibody Pair Kit you may need to optimize the assay. See ELISA Development and Optimization for more information. |
Wells scratched with pipette or washing tips | Use caution when dispensing and aspirating into and out of wells. Automated plate washers may need to be calibrated so tips don’t touch bottom of wells. |
Plate read at incorrect wavelength | Manufactured kits have optimized protocols. Make sure to use recommended wavelength/filter. Ensure plate reader is set accurately for type of substrate being used. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Plate sealers not used or reused | During incubations, cover assay plates with plate sealers. Use a fresh sealer each time the plate is opened. This will prevent wells from contaminating each other. |
Incorrect dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Longer incubation times than recommended | Manufactured kits have optimized protocols. Make sure to follow recommended incubation times. If developing ELISA using antibody pairs you may need to optimize the assay. See ELISA Development and Optimization for more information. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. Increasing duration of soak steps may also help. Add 30 seconds each time you let wash buffer soak. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Substrate exposed to light prior to use | Ensure substrate is not exposed to light—store in a dark place. Limit exposure to light while running assay. |
Longer incubation times than recommended | Manufactured kits have optimized protocols. Make sure to follow recommended incubation times. If developing ELISA using antibody pairs you may need to optimize the assay. See ELISA Development and Optimization for more information. |
Incorrect standard curve dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Possible Cause | Solution |
---|---|
Incorrect standard curve dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Capture antibody didn’t bind to plate | Ensure that you are using an ELISA plate, not a tissue culture plate. Dilute antibody in PBS. Ensure correct preparation and incubation time for both coating and blocking steps. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. Increasing duration of soak steps may also help. Add 30 seconds each time you let wash buffer soak. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Capture antibody didn’t bind to plate | Ensure that you are using an ELISA plate, not a tissue culture plate. Dilute antibody in PBS. Ensure correct preparation and incubation time for both coating and blocking steps. |
Plate sealers not used or reused | During incubations, cover assay plates with plate sealers. Use a fresh sealer each time the plate is opened. This will prevent wells from contaminating each other. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Inconsistent incubation temperature | Manufactured kits have optimized protocols. Make sure to follow recommended incubation temperatures. Be aware of fluctuations in temperature due to environmental conditions. |
Plate sealers not used or reused | During incubations, cover assay plates with plate sealers. Use a fresh sealer each time the plate is opened. This will prevent wells from contaminating each other. |
Incorrect dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Possible Cause | Solution |
---|---|
Uneven temperature | Seal the plate completely with a plate sealer during incubations. If 37oC incubation is indicated make sure plate is in the center of incubator. |
Evaporation | Seal the plate completely with a plate sealer during incubations. |
Stacked plates | Avoid stacking plates during incubation. |
Possible Cause | Solution |
---|---|
Reagents not at room temperature at start of assay | It is recommended that all reagents be at room temperature before starting the assay. Allow reagents to sit on bench for 15–20 minutes to reach room temperature. |
Incorrect storage of components | Double check storage conditions on kit label. Most kits need to be stored at 2–8oC. |
Expired reagents | Confirm expiration dates on all reagents. Do not use reagents that are past the expiration date. |
Reagents added/prepared incorrectly | Check protocol, ensure reagents were added in the proper order and prepared to correct dilution. |
Incorrect dilutions prepared | Check pipetting technique—see below—and double check calculations. |
Capture antibody didn’t bind to plate |
|
Not enough detector antibody used | Manufactured kits have optimized protocols. Make sure to follow recommended antibody dilutions. If developing ELISA using an Antibody Pair Kit you may need to optimize the assay. See ELISA Development and Optimization for more information. |
Wells scratched with pipette or washing tips | Use caution when dispensing and aspirating into and out of wells. Automated plate washers may need to be calibrated so tips don’t touch bottom of wells. |
Plate read at incorrect wavelength | Manufactured kits have optimized protocols. Make sure to use recommended wavelength/filter. Ensure plate reader is set accurately for type of substrate being used. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Plate sealers not used or reused | During incubations, cover assay plates with plate sealers. Use a fresh sealer each time the plate is opened. This will prevent wells from contaminating each other. |
Incorrect dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Longer incubation times than recommended | Manufactured kits have optimized protocols. Make sure to follow recommended incubation times. If developing ELISA using antibody pairs you may need to optimize the assay. See ELISA Development and Optimization for more information. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. Increasing duration of soak steps may also help. Add 30 seconds each time you let wash buffer soak. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Substrate exposed to light prior to use | Ensure substrate is not exposed to light—store in a dark place. Limit exposure to light while running assay. |
Longer incubation times than recommended | Manufactured kits have optimized protocols. Make sure to follow recommended incubation times. If developing ELISA using antibody pairs you may need to optimize the assay. See ELISA Development and Optimization for more information. |
Incorrect standard curve dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Possible Cause | Solution |
---|---|
Incorrect standard curve dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Capture antibody didn’t bind to plate | Ensure that you are using an ELISA plate, not a tissue culture plate. Dilute antibody in PBS. Ensure correct preparation and incubation time for both coating and blocking steps. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. Increasing duration of soak steps may also help. Add 30 seconds each time you let wash buffer soak. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Capture antibody didn’t bind to plate | Ensure that you are using an ELISA plate, not a tissue culture plate. Dilute antibody in PBS. Ensure correct preparation and incubation time for both coating and blocking steps. |
Plate sealers not used or reused | During incubations, cover assay plates with plate sealers. Use a fresh sealer each time the plate is opened. This will prevent wells from contaminating each other. |
Possible Cause | Solution |
---|---|
Insufficient washing | Use appropriate washing procedure—see below. At the end of each washing step, invert plate on absorbent tissue and allow to completely drain, tapping forcefully if necessary to remove any residual fluid. |
Inconsistent incubation temperature | Manufactured kits have optimized protocols. Make sure to follow recommended incubation temperatures. Be aware of fluctuations in temperature due to environmental conditions. |
Plate sealers not used or reused | During incubations, cover assay plates with plate sealers. Use a fresh sealer each time the plate is opened. This will prevent wells from contaminating each other. |
Incorrect dilutions prepared | Check pipetting technique—see below—and double-check calculations. |
Possible Cause | Solution |
---|---|
Uneven temperature | Seal the plate completely with a plate sealer during incubations. If 37oC incubation is indicated make sure plate is in the center of incubator. |
Evaporation | Seal the plate completely with a plate sealer during incubations. |
Stacked plates | Avoid stacking plates during incubation. |
This 72-page guide provides detailed information about different tools for protein and RNA quantitation. Download this valuable technical resource that covers technologies useful for cancer and inflammation research, immunology, neurology and more. Learn more about how antibody pairs, ELISA kits, and multiplex kits for the Invitrogen Luminex platform may help advance your research.
For Research Use Only. Not for use in diagnostic procedures.