Our slow-response potential-sensitive probes ref (see Figure 22.1.1B in Introduction to Potentiometric Probes—Section 22.1) are listed in Characteristics and selected applications of Molecular Probes slow-response probes—Table 22.2, along with their charges, optical responses and selected applications.

Carbocyanines

DiI, DiS and DiO Derivatives

Indo- (DiI), thia- (DiS) and oxa- (DiO) carbocyanines with short alkyl tails (<7 carbon atoms) were among the first potentiometric fluorescent probes developed.ref These cationic dyes accumulate on hyperpolarized membranes and are translocated into the lipid bilayer.ref Aggregation within the confined membrane interior usually results in decreased fluorescence, although the magnitude and even the direction of the fluorescence response is strongly dependent on the concentration of the dye and its structural characteristics.ref While the distribution of extracellularly applied dye is dependent on both the plasma and mitochondrial membrane potentials, the primary determinant is the latter. Very low applied concentrations (<100 nM) are required to obtain mitochondrial signal specificity and avoid potential-independent background derived from staining of the endoplasmic reticulum and other intracellular membranes.

DiOC6(3) (D273) has been the most widely used carbocyanine dye for membrane potential measurements,ref followed closely by DiOC5(3); see Characteristics and selected applications of Molecular Probes slow-response probes—Table 22.2 for selected references. In flow cytometry measurements, the detected intensity of carbocyanine fluorescence is dependent not only on the membrane potential, but also on cell size. In some cases, measurements of forward light scatter have been used to normalize the optical changes for cell size variability. A fluorescence ratio method (Figure 22.3.1) that exploits a potential-dependent red shift in the emission spectrum of DiOC2(3) (D14730) has been developed for membrane potential measurements in bacteria ref (B34950, see below). The 633 nm light–excitable indodicarbocyanine DiIC1(5) ref enables analysis of mitochondrial potential in apoptotic cells in combination with fluoresceinated annexin V ref (A13199, Assays for Apoptosis—Section 15.5), a method we have utilized in two of our MitoProbe Assay Kits for flow cytometry (M34150, M34151; see below). Carbocyanine dyes, particularly thiacyanines such as DiSC3(5) (D306), can inhibit respiration ref and may therefore be relatively cytotoxic.ref

slow-response-probes.par.0487.image.275.457.1.s000933-red-green-fluorescence-gif


Figure 22.3.1 A) Two-color flow cytometric analysis of Staphylococcus aureus populations stained with 30 µM DiOC2(3) (D14730) in the presence (red) or absence (blue) of the metabolic uncoupler carbonyl cyanide m-chlorophenylhydrazone (CCCP). Note the variability (~100-fold range) of the green and red fluorescence intensities. B) The same data expressed as red/green fluorescence intensity ratios. Ratio values are calculated by subtracting the logarithmic green fluorescence channel value from the corresponding logarithmic red fluorescence channel value.ref Figure supplied by Howard M. Shapiro, Harvard Medical School, Boston, MA.

JC-1 and JC-9

JC-1 (5,5',6,6'-tetrachloro-1,1',3,3'-tetraethylbenzimidazolylcarbocyanine iodide, T3168) exists as a green-fluorescent monomer at low concentrations or at low membrane potential. However, at higher concentrations (aqueous solutions above 0.1 µM) or higher potentials, JC-1 forms red-fluorescent "J-aggregates," which exhibit a broad excitation spectrum and a very narrow emission spectrum (photo). Because J-aggregate formation increases linearly with applied membrane potential over the range of 30–180 mV, this phenomenon can be exploited for potentiometric measurements ref (Characteristics and selected applications of Molecular Probes slow-response probes—Table 22.2). JC-1 is more specific for mitochondrial versus plasma membrane potential and more consistent in its response to depolarization than some other cationic dyes such as DiOC6(3) and rhodamine 123.ref

Various types of ratio measurements are possible by combining signals from the green-fluorescent JC-1 monomer (absorption/emission maxima ~514/529 nm) and the red-fluorescent J-aggregate (absorption/emission maxima ~585/590 nm), which can be effectively excited anywhere between 485 nm and its absorption maximum. Optical filters designed for fluorescein and tetramethylrhodamine can be used to separately visualize the monomer and J-aggregate forms, respectively. Alternatively, both forms can be observed simultaneously using a fluorescein longpass optical filter set. For flow cytometry, JC-1 can be excited at 488 nm and detected in bivariate mode using the green channel for the monomer and the red channel for the J-aggregate form ref (Figure 22.3.2).

JC-1 is widely used for detecting mitochondrial depolarization in apoptotic cells ref (MitoProbe JC-1 Assay Kit, M34152, described below) and for assaying multidrug-resistant cells ref (Probes for Cell Adhesion, Chemotaxis, Multidrug Resistance and Glutathione—Section 15.6). It is also frequently employed for mitochondrial function assessment in cell-based high-throughput assays.ref We have discovered another carbocyanine dye, JC-9 (3,3'-dimethyl-α-naphthoxacarbocyanine iodide, D22421), with potential-dependent spectroscopic properties (photo) similar to those of JC-1 for detecting mitochondrial depolarization in apoptotic cells.ref

slow-response-probes.par.82010.image.650.485.1.s000932-mitochondrial-gif
Figure 22.3.2 Bivariate JC-1 (T3168) analysis of mitochondrial membrane potential in HL60 cells by flow cytometry. The sensitivity of this technique is demonstrated by the response to depolarization using K+/valinomycin (V1644, Fluorescent Na+ and K+ Indicators—Section 21.1) (bottom two panels). Distinct populations of cells with different extents of mitochondrial depolarization are detectable following apoptosis-inducing treatment with 5 µM staurosporine for two hours (top right panel). Figure courtesy of Andrea Cossarizza, University of Modena and Reggio Emilia, Italy.

MitoProbe JC-1 Assay Kit for Flow Cytometry

The MitoProbe JC-1 Assay Kit (M34152) provides the cationic dye JC-1 and a mitochondrial membrane potential disrupter, CCCP (carbonyl cyanide 3-chlorophenylhydrazone), for the study of mitochondrial membrane potential. Use of JC-1 fluorescence ratio detection allows researchers to make comparative measurements of membrane potential and to determine the percentage of mitochondria within a population that respond to an applied stimulus (Figure 22.3.3). Subtle heterogeneity in cellular responses can be discerned in this way.ref For example, four distinct patterns of mitochondrial membrane potential change in response to glutamate receptor activation in neurons have been identified using confocal ratio imaging of JC-1 fluorescence.ref

Each MitoProbe JC-1 Assay Kit provides:

Sufficient reagents are provided for 100 assays, based on a labeling volume of 1 mL.

slow-response-probes.par.14379.image.275.457.1.s002435-flow-cytometric-gif


Figure 22.3.3 Flow cytometric analysis of Jurkat cells using the MitoProbe JC-1 Assay Kit (M34152). Jurkat cells were stained with 2 µM JC-1 for 15 minutes at 37°C, 5% CO2, and then washed with phosphate-buffered saline (PBS) and analyzed on a flow cytometer using 488 nm excitation with 530 nm and 585 nm bandpass emission filters. Untreated cultured cells are shown in panel A. Panel B shows cells induced to apoptosis with 10 µM camptothecin for 4 hours at 37°C.

MitoProbe DiIC1(5) and MitoProbe DiOC2(3) Assay Kits for Flow Cytometry

The MitoProbe DiIC1(5) and MitoProbe DiOC2(3) Assay Kits (M34151, M34150) provide solutions of the far-red–fluorescent DiIC1(5) (1,1',3,3,3',3'-hexamethylindodicarbocyanine iodide) and green-fluorescent DiOC2(3) (3,3'-diethyloxacarbocyanine iodide) carbocyanine dyes, respectively, along with a mitochondrial membrane potential uncoupler, CCCP, for the study of mitochondrial membrane potential. These DiIC1(5) and DiOC2(3) carbocyanine dyes penetrate the cytosol of eukaryotic cells and, at concentrations below 100 nM, accumulate primarily in mitochondria with active membrane potentials. In the case of DiOC2(3), this accumulation is accompanied by a shift from green to red emission due to dye stacking (Figure 22.3.1), allowing the use of a ratiometric parameter (red/green fluorescence ratio) that corrects for size differences when measuring membrane potential in bacteria.ref DiOC2(3) can be paired with other reagents, such as the far-red–fluorescent allophycocyanin annexin V (A35110, Assays for Apoptosis—Section 15.5) or TO-PRO-3 nucleic acid stain (T3605, Nucleic Acid Stains—Section 8.1) for multiparameter study of vitality and apoptosis ref (Figure 22.3.4).

The MitoProbe DiIC1(5) and MitoProbe DiOC2(3) Assay Kits provide:

Each kit provides sufficient reagents for 100 assays, based on a labeling volume of 1 mL.

slow-response-probes.par.45519.image.275.246.1.s002431-flow-cytometric-jurkat-gif


Figure 22.3.4 Flow cytometric analysis of camptothecin-treated Jurkat cells stained with DiOC2(3) (D14730, M34150) and allophycocyanin annexin V (A35110). Jurkat cells were incubated for 4 hours with camptothecin at 37°C, 5% CO2, then stained with 50 nM DiOC2(3) and allophycocyanin annexin V. Cells were analyzed on a flow cytometer using 488 nm and 633 nm excitations with 530 nm and 660 nm bandpass emission filters.

BacLight Bacterial Membrane Potential Kit

The BacLight Bacterial Membrane Potential Kit (B34950) provides a fluorescent membrane-potential indicator dye, DiOC2(3), along with a proton ionophore (CCCP) and premixed buffer. At low concentrations, DiOC2(3) exhibits green fluorescence in all bacterial cells, but it becomes more concentrated in healthy cells that are maintaining a membrane potential, causing the dye to self-associate and the fluorescence emission to shift to red. The red- and green-fluorescent bacterial populations are easily distinguished using a flow cytometer (Figure 22.3.5). CCCP is included in the kit for use as a control because it eradicates the proton gradient, eliminating bacterial membrane potential.ref

Each BacLight Bacterial Membrane Potential Kit contains:

Using the recommended reagent dilutions and volumes, this kit provides sufficient DiOC2(3) to perform approximately 100 individual assays by flow cytometry; sufficient CCCP is provided for 30 depolarized control samples. The BacLight Bacterial Membrane Potential Kit is designed to assay bacterial concentrations between 105 and 107 organisms per mL. Note that DiOC2(3) and CCCP are inhibitors of respiration, rendering the cells nonculturable beyond the brief time window required for staining and analysis.

Using the BacLight Bacterial Membrane Potential Kit, we have detected membrane potentials in all bacteria tested (including logarithmically growing cultures of Micrococcus luteus, Staphylococcus aureus, Bacillus cereus, Staphylococcus warnerii, Escherichia coli and Salmonella choleraesuis), although the magnitude varies with species (Figure 22.3.6).

slow-response-probes.par.81539.image.275.209.1.s002297-staphylococcus-gif


Figure 22.3.5 Response of Staphylococcus aureus to valinomycin and external potassium ions, as measured by flow cytometry using the BacLight Bacterial Membrane Potential Kit (B34950). Samples containing S. aureus were treated with 5 µM valinomycin in different concentrations of potassium buffer, and then stained using 30 µM DiOC2(3) for 30 minutes, according to the kit protocol. Data are expressed either using a ratiometric parameter based on the formula provided in the kit protocol (triangles, right axis) or as the ratio of population mean red-fluorescence intensity/mean green-fluorescence intensity (circles, left axis).

slow-response-probes.par.1632.image.275.307.1.s002296-baclight-membrane-gif


Figure 22.3.6 Detection of membrane potential in various bacteria with the BacLight Bacterial Membrane Potential Kit (B34950). Red/green fluorescence ratios were calculated using population mean fluorescence intensities for gram-positive (Micrococcus luteus, Staphylococcus aureus, Bacillus cereus and Staphylococcus warnerii) and gram-negative (Escherichia coli and Salmonella choleraesuis) bacteria incubated with 30 µM DiOC2(3) for 30 minutes in either the presence or absence of 5 µM CCCP, according to the kit protocol.

Rhodamine 123, TMRM and TMRE

Rhodamine 123 (R302) is widely used as a structural marker for mitochondria (Probes for Mitochondria—Section 12.2) and as an indicator of mitochondrial activity (Viability and Cytotoxicity Assay Reagents—Section 15.2, photo). Highly selective, potential-dependent staining of mitochondria is obtained by setting the extracellular K+ concentration close to intracellular values (~137 mM), thereby depolarizing the plasma membrane.ref

TMRM (T668) and TMRE (T669), the methyl and ethyl esters of tetramethylrhodamine are closely related to rhodamine 123. They are primarily mitochondrial membrane potential sensors.ref As with rhodamine 123, accumulation of these cationic dyes in mitochondria results in diminished fluorescence due to self-quenching ref (photo). TMRM and TMRE cross the plasma membrane more rapidly than rhodamine 123, and their strong fluorescence allows the use of low probe concentrations, thus avoiding aggregation. Because their fluorescence is relatively insensitive to the environment, spatially resolved fluorescence of TMRM and TMRE presents an unbiased profile of their transmembrane distribution that can be directly related to the plasma membrane potential via the Nernst equation.ref TMRE has been successfully used in a high-throughput screening assay for drugs that affect mitochondrial membrane potential in live cells.ref TMRM has been used in conjunction with X-rhod-1 AM (X14210, Fluorescent Ca2+ Indicators Excited with Visible Light—Section 19.3) for simultaneous confocal imaging of mitochondrial membrane potential and calcium in rat cardiomyocytes.ref

Oxonols

Oxonol V and Oxonol VI

The anionic bis-isoxazolone oxonols accumulate in the cytoplasm of depolarized cells by a Nernst equilibrium–dependent uptake from the extracellular solution.ref Their voltage-dependent partitioning between water and membranes is often measured by absorption rather than fluorescence. Of the oxonols studied by Smith and Chance,ref oxonol VI gave the largest spectral shifts, with an isosbestic point at 603 nm. In addition, oxonol VI responds to changes in potential more rapidly than oxonol V and is therefore considered to be the better probe for measuring fast potential changes.ref

DiBAC (Bis-Oxonol) Dyes

The three bis-barbituric acid oxonols, often referred to as DiBAC dyes, form a family of spectrally distinct potentiometric probes with excitation maxima at approximately 490 nm (DiBAC4(3); B438), 530 nm (DiSBAC2(3), B413) and 590 nm (DiBAC4(5)). Several papers have referred to these dyes simply as "bis-oxonol" and it is not always possible to determine which of the dyes was employed; however, DiBAC4(3) has been used in the majority of publications that cite a "bis-oxonol."

These dyes enter depolarized cells where they bind to intracellular proteins or membranes and exhibit enhanced fluorescence and red spectral shifts.ref Increased depolarization results in more influx of the anionic dye and thus an increase in fluorescence (photo). Conversely, hyperpolarization is indicated by a decrease in fluorescence (Figure 22.3.7). In contrast to cationic carbocyanines, anionic bis-oxonols are largely excluded from mitochondria and are primarily sensitive to plasma membrane potential. Potential-dependent fluorescence changes generated by DiBAC4(3) are typically ~1% per mV.ref Interactions between anionic oxonols and the cationic K+-valinomycin complex complicate the use of this ionophore when calibrating potentiometric responses.ref Oxonol dyes have known pharmacological activity against various ion channels and receptors.ref It is therefore important to establish, as is the case in any experiment using fluorescent probes, that experimental observations with implied physiological significance are independent of the externally applied probe concentration.

slow-response-probes.par.77673.image.275.240.1.s001683-atp-sensitive-potassium-gif


Figure 22.3.7 Detection of ATP-sensitive potassium (KATP) channel activation in isolated capillaries from guinea pig hearts using DiBAC4(3) (B438), a slow potentiometric probe. Application of a K+ channel opener (HOE 234) induced membrane hyperpolarization, resulting in a net efflux of intracellular DiBAC4(3), which is registered as a decrease of fluorescence intensity. These effects were reversed by subsequent treatment with the channel blocker glibenclamide. Figure reproduced with permission from ref.

Merocyanine 540

Although merocyanine 540 was among the first fluorescent dyes to be used as a potentiometric probe,ref its use for this application has declined with the advent of superior probes. A significant disadvantage of merocyanine 540 is its extreme phototoxicity; consequently, it is now more commonly used as a photosensitizer.ref

Merocyanine 540 exhibits a biphasic kinetic response to membrane polarization changes. It binds to the surface of polarized membranes in a perpendicular orientation, reorienting as the membrane depolarizes to form nonfluorescent dimers with altered absorption spectra.ref This fast (microseconds) reorientation is followed by a slower response caused by an increased dye uptake.

Merocyanine 540 is also a useful probe of lipid packing because it binds preferentially to membranes with highly disordered lipids.ref Fluorescence of merocyanine 540 is sensitive to heat-induced changes in the organization of membrane lipids.ref

Data Table

For a detailed explanation of column headings, see Definitions of Data Table Contents

Cat. No.
MWStorageSolubleAbsECEmSolventNotes
B413
DiSBAC2(3)
436.54LDMSO, EtOH535170,000560MeOH1
DiBAC4(5)542.67LDMSO, EtOH590160,000616MeOH1
B438
DiBAC4(3)
516.64LDMSO, EtOH493140,000516MeOH1, 2
D273
DiOC6(3)
572.53D,LDMSO484154,000501MeOH 
D306
DiSC3(5)
546.53D,LDMSO651258,000675MeOH 
D14730
DiOC2(3)
460.31D,LDMSO482165,000497MeOH3
D22421
JC-9 (DiNOC1(3))
532.38D,LDMSO, DMF522143,000535CHCl34
DiIC1(5)510.46LDMSO638230,000658MeOH5
merocyanine 540569.67D,LDMSO, EtOH555143,000578MeOH 
oxonol V384.39LDMSO, EtOH610135,000639MeOH1
oxonol VI316.36LDMSO, EtOH599136,000634MeOH1
R302
rhodamine 123
380.83F,D,LMeOH, DMF507101,000529MeOH 
T668
tetramethylrhodamine, methyl ester (TMRM)
500.93F,D,LDMSO, MeOH549115,000573MeOH 
T669
tetramethylrhodamine, ethyl ester (TMRE)
514.96F,D,LDMSO, EtOH549109,000574MeOH 
T3168
JC-1 (CBIC2(3))
652.23D,LDMSO, DMF514195,000529MeOH6
  1. Oxonols may require addition of a base to be soluble.
  2. Fluorescence of DiBAC4(3) increases about 3-fold relative to H2O on binding to proteins (Abs = 499 nm, Em = 519 nm).ref
  3. QY for DiOC2(3) in MeOH = 0.04. Abs = 478 nm, Em = 496 nm in H2O.ref
  4. JC-9 exhibits long-wavelength J-aggregate emission at ~635 nm in aqueous solutions and polarized mitochondria.
  5. DiIC1(5) in H2O; Abs = 636 nm, Em = 658 nm.
  6. JC-1 forms J-aggregates with Abs/Em = 585/590 nm at concentrations above 0.1 µM in aqueous solutions (pH 8.0).ref

For Research Use Only. Not for use in diagnostic procedures.