Search Thermo Fisher Scientific
Search Thermo Fisher Scientific
We are dedicated to your success. Get back on track. View our expert recommendations for commonly encountered problem scenarios.
View the relevant questions below:
Beginning your experiment? Visit our
Here are some possible causes and solutions for this problem:
Cause | Solution |
Placing the plate on an absorbent material both at the bench and on the shaker (which results in wicking when there is contact between the absorbent material and the opening on the bottom of the well funnel).
| Rest the plate currently in use on top of a second filter plate so that the filter plate currently in use only comes into contact with a nonabsorbent surface. This will also help locate any potential leaks. Prior to incubations, use a Kimwipe™ tissue and lightly press up on each well to dry off the bottom of the plate. |
Putting any vertical pressure on the plate during the incubations (using either tape or clamps). | Use clamps that fit around the sides of the filter plate to secure the plate to the shaker during incubations. |
Subjecting the plate to vacuum pressure greater than 5 mm Hg, even for a moment, which either tears the filter membrane or creates an opening that exceeds the design limits for volume retention. | Prevent any vacuum surge by opening and adjusting the vacuum on the manifold before placing the plate on the manifold surface and checking the vacuum with a test plate (i.e., not the plate used for the assay). For our assays we recommend the setting not exceed 5 mm Hg. |
Puncturing the filter membrane with pipet tips, by inserting the tips all the way into the wells during reagent loading. | Pipet solutions along the sides of the wells, rather than deep into the wells. |
Separating the soft (opaque) plastic bottom from the hard (clear) plastic top of the filter plate before starting, which compromises the integrity of the seal even if the pieces appear to fit back together properly. | If you accidentally separate these two layers of the plate, the plate should be discarded and a fresh plate used instead. |
This may originate from the probe height being set too low, and the pressure from the fluid return is forcing liquid through the membrane. However, remember that as long as the required 100 beads are being read in a reasonable time frame, the sample is being read correctly. If there is difficulty reaching 100 beads, liquid may have leaked out prior to reading the sample. In this case, stop the run and check the plate for leakage. If leakage is found, remove the Wash Solution on the vacuum manifold and completely dry the bottom of the plate. Add fresh Wash Solution and shake. Continue the run from where it was stopped.
Note: Refer to the appropriate instrument hardware manual for instructions to check and to reset the sample probe (sample needle) height.
If wells are clogged after the first incubation with the standards:
Here are some suggestions:
Here are some suggestions:
This pattern is indicative of a sample matrix effect. Here are some suggestions:
Here are some suggestions:
This indicates that an incorrect buffer was used for the final step. The wash solution provided in the kit must be used for washing the beads and for resuspending the beads before loading them into the Luminex™ instrument. The osmolarity of the solution will impact the size of the bead, and any change in the bead size will alter detection by the instrument.
Here are possible causes and solutions for this issue:
Here are possible causes and solutions for this issue:
For Research Use Only. Not for use in diagnostic procedures.