Search Thermo Fisher Scientific
Physical Characteristics | Instrument Type: | Benchtop device for imaging cells |
Instrument Dimensions: | 15.9" (W) × 21.1" (H) × 13.9" (D) 40.4 cm (W) × 53.6 cm (H) × 35.3 cm (D) | |
Weight: | 26 lbs (11.8 kg) | |
Operating Power: | 100–240 VAC | |
Frequency: | 50–60 Hz | |
Electrical input: | 5 VDC, 4.15 A | |
Operating Temperature: | 4–32°C | |
Operating Humidity: | <90% (non-condensing) | |
Output Ports: | 4 USB | |
Optics | Objective: | 20X plan fluorite; |
Numerical aperture (NA): | 0.45; | |
Working distance (WD): | 5.9 mm (defined as the tip of the objective to the bottom of a 1.2-mm thick slide) | |
Overall Magnification: | 460X (optical)–1840X (with digital zoom) | |
Display Resolution: | 1366 × 768 pixels | |
Image Resolution: | 1296 × 964 pixels | |
Contrast Methods: | Fluorescence and transmitted light (relief phase) | |
Color Channels: | 4 channels (relief phase, blue fluorescence, green fluorescence, red fluorescence) | |
Illumination: | LED (50,000 hour life), adjustable intensity | |
Excitation: | Blue channel: 390/40 nm; Green channel: 482/18 nm; Red channel: 586/15 nm | |
Emission: | Blue channel: 446/33 nm; Green channel: 532/59 nm; Red channel: 646/68 nm | |
Stage to Condenser: | 60 mm | |
Camera: | Sony 1.3MP 1/3" ICX445 EXview HAD CCD | |
Hardware/Software | Stage: | Mechanical XY glide stage that can move within a 5 mm diameter circle |
LCD monitor: | 15-inch color; display resolution 1366 x 768 pixels; Image resolution 1296 x 964 pixels; adjustable tilt | |
Software: | FLoid® Cell Imaging Station Software | |
Captured Images: | 16-bit monochochrome TIFF, PNG, BMP and JPG; 1296 × 964 pixels | |
Supplied USB Drive: | 2 Gigabyte |