Charging effects in X-ray photoelectron spectroscopy

When X-ray photons strike a surface, they cause the emission of electrons. This, of course, is the basis of the XPS technique. If the surface is electrically insulating, then the emission of electrons causes a positive charge to accumulate at the surface. When a non-monochromated source is used for XPS analysis there are usually a sufficient number of electrons in the region of the sample to limit the effect of the charging. However, when using a monochromated X-ray source for XPS measurements, there are too few stray electrons to control the charging and the resulting positive charge severely affects the XPS spectrum. It causes the peaks in the spectrum to shift to high binding energies and become distorted. For this reason, it is necessary to neutralize the charge on the surface by replenishing electrons from an external source. This stabilizes and controls the charging to within a few electron volts of the neutral state. Any remaining small shift can be removed after data collection is complete, by using the spectrum processing tools contained within the Thermo Scientific Avantage data system.


Combination electron beam and ion beam gun

The use of a single gun to produce both the focused low energy electron beam and the large area ion flux has a number of distinct advantages over the use of independent sources. A new, combined ion/electron gun has been designed which incorporates all the above advantages. This novel design uses a LaB6 emitter to produce a bright source of low energy electrons, with an energy spread of only about 0.3 eV. These are focused to a small spot on the sample. Precise alignment to the X-ray beam is provided using integral X-Y deflectors. Focusing and alignment of this low energy beam is enhanced by the extremely low residual magnetic fields present in the mu-metal analysis chambers.

Features of Charge Compensation for the ESCALAB Xi+ spectrometer

  • Patented ‘in the lens’ multi-mode electrostatic flood source
  • 0 - 5 eV (mode 1) charge compensation
  • 0 - 1000 eV (mode 2) REELS / imaging / alignment
  • Up to 500 µA electron emission
  • Additional dual beam (ion / electron) flood source

Features of Charge Compensation for Nexsa and K-Alpha spectrometers

  • Patented dual beam source
  • Single mode flood gun for all sample types
  • Single mode for all X-ray spot sizes
  • Ultra-low energy beam results in minimal shift from reference positions
  • Flood source assembly incorporating electrostatic deflection for precise alignment
  • REELS capability available for Nexsa
Charging Effects in XPS

Products

Style Sheet to change H2 style to p with em-h2-header class
Style Sheet for Instrument Cards Origin

Nexsa G2 XPS

  • マイクロフォーカスX線源
  • 独自の追加可能なオプション
  • 単原子およびクラスターイオンデプスプロファイリング用のデュアルモードイオン源

K-Alpha XPS

  • 高分解能XPS
  • 高速で効率的な自動ワークフロー
  • イオン源を用いたデプスプロファイリング

ESCALAB Xi+ XPS

  • 高いスペクトル分解能
  • マルチテクニックによる表面分析
  • 試料作製および拡張のための豊富なオプション

Contact us

Style Sheet for Support and Service footer
Style Sheet for Fonts
Style Sheet for Cards

材料科学向けの
電子顕微鏡サービス

最適なシステム性能をお届けするため、当社は国際的なネットワークで、分野ごとのサービスエキスパート、テクニカルサポート、正規交換部品などを提供しています。